Measuring range: 0-100%
Output signal in air: 9-14mV
Circuit interface: 3Pin Molex connector
Accuracy and repeatability: <1% volume oxygen, after calibration in 100% oxygen
Linearity error: <3% relative error
Response time: <12 seconds relative to 90% relative
Zero drift: <200uV in 100% nitrogen for 5 minutes
Cross Interference: <0.5% volume oxygen for:
10% CO2, in equilibrium nitrogen
80%N2O, in balance nitrogen
7.5% Halothane in equilibrated nitrogen
7.5% isoflurane in equilibrated nitrogen
7.5% enflurane in equilibrium nitrogen
9% enflurane in equilibrated nitrogen
20% desflurane in equilibrated nitrogen
Effect of temperature: -0.03% relative temperature, 25°C
Effect of Pressure: Proportional to partial pressure of oxygen
Effect of mechanical shock: <1% relative error, after falling from a height of 1 meter
Working temperature: 0℃-50℃
Temperature Compensation: Built-in thermistor compensation
Temperature influence: 25℃-40℃, 3% relative error
Compensation (steady state): Between 0°C and 50°C, 8% relative error
Working humidity: 0-99%RH (can be condensed in air for several hours)
Storage temperature: -20°C -50°C
Recommended storage temperature: 5℃-15℃
Recommended load: ≥10Ω
Warm-up time: <30 minutes after sensor replacement
Rated sensor life: ≥1 000 000% volume oxygen hours
Weight: about 28G